Recall Z-2948-2024 ยท BRY Cabinet, Table And Tray, Anesthesia
CareFusion 303, Inc.
BKIT BD Pyxis QFill Replenishment Station, REF: 155197-01, medication cabinet system
| Date | Aug 30, 2024 (initiated Jul 9, 2024) |
| Status | Open, Classified |
| Reason | Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label. |
| Root cause | Software design |
| Action | On July 11, 2024, CareFusion a subsidiary of Becton Dickinson (BD) issued a "Urgent Medical Device Correction" Notification to affected customers via Email. BD also notified customers through customer software release notes beginning June 24, 2024. BD asked consignees to take the following actions: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. 2. Users should verify the contents of the bin location being restocked. Actions to be Taken by the Customer: 3.Pl |
| Distribution | US Nationwide distribution in the states of CA, FL, KS, MN, NJ, NY, PA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.