Recall Z-2980-2024 ยท CDD Radioassay, Vitamin B12

Beckman Coulter Biomedical GmbH

PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.

DateSep 5, 2024 (initiated Jul 17, 2024)
StatusOpen, Classified
ReasonThe adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Root causeNonconforming Material/Component
ActionOn July 18, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. follow your laboratory procedure to handle biohazard materials and contact your Beckman Coulter representative for servicing your Power Express system. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of thi
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140496Access Vitamin B12 Assay, Access Ferritin Assay, Access Folate Assay, Access HYPERsensitive hTSH Assay, Power Express Sa
Beckman Coulter, Inc., Sep 16, 2014