ClearanceBrief
Recall Z-0082-2025 ยท BSZ Gas-Machine, Anesthesia

Datex-Ohmeda, Inc.

Aisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.

DateOct 18, 2024 (initiated Jul 12, 2024)
StatusOpen, Classified
ReasonPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Root causeDevice Design
ActionGE Healthcare notified consignees on about 07/12/2024 via letter. Consignees were instructed to ensure all potential users complete the Preoperative Checkout procedure in its entirety, as detailed in the URM, make sure that the breathing circuit is correctly connected, only connect an auxiliary manual breathing circuit to the ACGO port, place the included addendum with the URM and inspect all potentially affected devices for the presence of covers applied to the ACGO Port and ACGO Switch. Additionally, customers were requested to complete and return the response form.
Quantity15,739 units
DistributionWorldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Ca
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View, Av
Datex-Ohmeda, Inc., Nov 3, 2017