Recall Z-2416-2024 ยท RIB Microscopy Laser Instrument Or Accessory Products

EVIDENT SCIENTIFIC INC

SlideView VS200 slide viewer,

DateJul 26, 2024 (initiated Jul 22, 2024)
StatusOpen, Classified
ReasonA broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Root causeRadiation Control for Health and Safety Act
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.