Recall Z-2416-2024 ยท RIB Microscopy Laser Instrument Or Accessory Products
EVIDENT SCIENTIFIC INC
SlideView VS200 slide viewer,
| Date | Jul 26, 2024 (initiated Jul 22, 2024) |
| Status | Open, Classified |
| Reason | A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated. |
| Root cause | Radiation Control for Health and Safety Act |
| Distribution | US |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.