Recall Z-2974-2024 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Smiths Medical ASD Inc.

Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX

DateSep 11, 2024 (initiated May 31, 2024)
StatusOpen, Classified
ReasonThere is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
Root causeUnder Investigation by firm
ActionSmiths Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/31/2024 by letter delivered via traceable means. The notice explained the issue, potential risk, and requested the following: "Customer Required Actions: 1. Identify all affected paraPac plus serial numbers in your possession. 2. Perform an inspection to determine if your devices are affected per the instructions below: a. Rotate the tidal volume knob to the lowest position of the control and evaluate whether or not the knob stays in position as set or if it moves to a higher position. b. Repeat this
Quantity10,240 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123957PNEUPAC PARAPAC PLUS
Smiths Medical International , Ltd., Sep 16, 2013