Recall Z-2973-2024 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Smiths Medical ASD Inc.

Pneupac paraPAC plus 300 ventilator kit, REF P300NXX

DateSep 11, 2024 (initiated May 31, 2024)
StatusOpen, Classified
ReasonThere is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
Root causeUnder Investigation by firm
ActionSmiths Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/31/2024 by letter delivered via traceable means. The notice explained the issue, potential risk, and requested the following: "Customer Required Actions: 1. Identify all affected paraPac plus serial numbers in your possession. 2. Perform an inspection to determine if your devices are affected per the instructions below: a. Rotate the tidal volume knob to the lowest position of the control and evaluate whether or not the knob stays in position as set or if it moves to a higher position. b. Repeat this
Quantity2198 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123957PNEUPAC PARAPAC PLUS
Smiths Medical International , Ltd., Sep 16, 2013