Recall Z-2686-2024 ยท MJA Susceptibility Test Powders, Antimycobacterial

Becton Dickinson & Co.

BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA) Catalog Number: 245128

DateAug 16, 2024 (initiated Jul 18, 2024)
StatusOpen, Classified
ReasonMay intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.
Root causeUnder Investigation by firm
Action***UPDATE 11/13/2025*** BD Issued an URGENT: Medical Device Correction letters were sent to customers. BD is pleased to announce that we have resumed production of a modified version of the BD BACTEC" MGIT" 960 PZA Kit which includes modified inoculation methods and reduced shelf-life. BD has conducted a thorough review of the performance from all supported inoculum sources using the synthetic raw material and the following change has been implemented for the BD BACTEC" MGIT" 960 PZA Kit. 1. Only inoculum prepared from MGIT tubes 3-5 days post-positivity can be currently supported. 2. Inoculum
Quantity12,383 eaches
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brunei, Brazil, Canada, China, Costa Rica, Dominican Republic, Guatemala, Guyana, Hong Kong, Haiti, Indonesia, India, Japan, Korea, Myanmar, Macau, Mexico, Malaysia, Nepal, New Zealand, Panama, Per
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021582BACTEC MGIT 960 PZA KIT
Becton, Dickinson & CO, Jul 13, 2002