Recall Z-2413-2024 ยท OAS X-Ray, Tomography, Computed, Dental

Palodex Group Oy

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

DateAug 12, 2024 (initiated Oct 28, 2021)
StatusOpen, Classified
ReasonIt was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
Root causeRadiation Control for Health and Safety Act
ActionPalodex Group Oy is preparing a Class IlIl field correction to update the firmware of ORTHOPANTOMOGRAPH OP 3D Pro of the 27 defective units and these devices will be retested for Pan ADC reproducibility. All affected persons will be notified of this defect in accordance with 21 CFR 806.10.
Quantity103
DistributionUS Nationwide-Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163423Orthopantomograph OP300
Palodex Group OY, Aug 31, 2017