Recall Z-2514-2024 ยท PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System

BioFire Diagnostics, LLC

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and iden

DateAug 6, 2024 (initiated May 16, 2024)
StatusOpen, Classified
ReasonDue to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
Root causeUnder Investigation by firm
ActionOn 07/11/2024, the firm emailed an "URGENT FIELD SAFETY NOTICE" to customers informing them the firm has identified an increased risk of false positive Vibrio/Vibrio Cholerae results when the BIOFIRE GI Panel is used with certain lots of Remel" Cary-Blair transport medium (see Appendix A - Affected Remel" Cary-Blair Lots). Customers are instructed to: -If the BIOFIRE GI Panel is used in conjunction with certain lots of Remel" Cary-Blair (Lot # 743930, 769240, 712258, 782425, 732357, and 742049), positive results for Vibrio/Vibrio Cholerae should be confirmed by another method prior to reporti
Quantity67,613 kits (or 2,028,309 pouches). Each pouch is used for each individual test.
DistributionU.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140407FILMARRAY GASTROINTESTINAL (GI) PANEL
Biofire Diagnostics, May 2, 2014
K143005FilmArray Gastrointestinal (GI) Panel for use with the FilmArray 2.0
Biofire Diagnostics, LLC, Feb 19, 2015
K160459FilmArray Gastrointestinal Panel (GI) for use with FilmArray Torch
Biofire Diagnostics, LLC, Apr 1, 2016
K230404BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
Biofire Diagnostics, LLC, Mar 16, 2023