Recall Z-2449-2024 ยท FRN Pump, Infusion
ICU Medical Inc
Plum 360/A+ spare batteries. Item Number: SUB0000864.
| Date | Aug 9, 2024 (initiated May 7, 2024) |
| Status | Open, Classified |
| Reason | Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected. |
| Root cause | Process control |
| Action | Consignees were notified on about 05/07/2024 via traceable mail with an "UPDATED URGENT MEDICAL DEVICE CORRECTION" letter to the original recall was expanding. Consignees were instructed to replace batteries the first time the pump issues a replace battery alarm. Continued use of pumps on battery after the first instance of the replace battery alarm will reduce the time from a low battery alarm to a depleted battery/infusion stopping. Additionally, pumps should be kept plugged into AC power, closely monitor if using battery power, have a backup pump available, and do not use a replacement ba |
| Quantity | 130,826 units |
| Distribution | Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay. |
| FDA record | Open on fda.gov |