Recall Z-2718-2024 ยท HRS Plate, Fixation, Bone

Treace Medical Concepts, Inc.

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

DateAug 23, 2024 (initiated Jun 2, 2024)
StatusOpen, Classified
ReasonPotential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.
Root causeUnder Investigation by firm
ActionDistributors were contacted via phone or in person to hold product starting on 06/02/2024. Recall notifications were disseminated on 06/13/2024 via email containing the recall letter, acknowledgement, and shipping label for return. On 07/29/2024, Treace Medical contacted surgeons via letter. Surgeons were instructed to monitor patients for adverse events and to complete and return the acknowledgement form.
Quantity40 kits containing 40 plates
DistributionUS Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K240173Treace Medical Systems (TMC) Plating System
Treace Medical Concepts, Inc., Feb 22, 2024