ClearanceBrief
Recall Z-3090-2024 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Smiths Medical ASD Inc.

paraPAC plus 310 ventilator kit with internal PEEP and CPAP, Item Number P310N

DateSep 20, 2024 (initiated May 31, 2024)
StatusOpen, Classified
ReasonSmiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
Root causeOther
ActionSmiths Medical issued an URGENT MEDIC AL DEVICE CORRECTION notice to its consignees on05/31/2024 via letter. The notice explained the issue, potential risk, and requested the following: "Perform an inspection to determine if your devices are affected, per the instructions below: a. First perform a visual inspection to determine if the outlet connector is disconnected. b. Second evaluate the connector physically to determine if the outlet connector is loose or moves when placing a patient circuit on the connector or when removing it. c. If outlet connector remains tight after physical inspec
Quantity11244 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123957PNEUPAC PARAPAC PLUS
Smiths Medical International , Ltd., Sep 16, 2013