Recall Z-2945-2024 ยท PRL Iliac Covered Stent, Arterial

W L Gore & Associates, Inc.

Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A BXB051502A BXB051901A BXB051902A BXB052901A BXB052902A BXB053901A BXB053902A BXB055901A BXB055902A BXB057901A BXB057902A BXB061501A BXB061502A BXB061901A BXB061902A BXB062901A BXB062902A BXB063901A BXB063902A BXB065901A BXB065902A BXB067901A BXB067902A BXB071501A BXB071502A

DateAug 30, 2024 (initiated Jul 1, 2024)
StatusOpen, Classified
ReasonDue to packaging defects, sterility assurance and heparin activity may be compromised.
Root causePackage design/selection
ActionBetween the dates of July 18-19,2024 , W.L. Gore issued a "Urgent Medical Device Correction" notification to affected consignees either via UPS or hand delivery. W.L. Gore asked consignees to take the following actions: 1. For US Only, per the IFU, carefully inspect the foil pouch to ensure that it is not compromised. Do not use the GORE VIABAHN VBX Balloon Expandable Endoprosthesis if the foil pouch is compromised. Do not use the GORE VIABAHN VBX Balloon Expandable Endoprosthesis if the sterile pouch is compromised. This warning includes the Compliance Chart embedded in the seal or a damage
Quantity42,901 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.