Recall Z-2543-2024 ยท DRM Compressor, Cardiac, External

Defibtech, LLC

RMU-2000 Automated Chest Compression Device

DateAug 14, 2024 (initiated Jul 12, 2024)
StatusOpen, Classified
ReasonProblem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.
Root causeProcess control
ActionOn July 12, 2024, the firm notified affected users via Urgent Medical Device Safety Removal letters. Affected units (as identified by serial number) should be quarantined by the customer. Defibtech will contact customers to arrange for the return of the product. Customers' devices will be refunded, repaired, or replaced at no cost to them. Customers who have further transferred the device to a downstream account must notify those accounts of this notice or provide Defibtech with the contact information for those sub-accounts. Defibtech Customer Support: 1-877-453-4507, 7:30AM - 6PM Eastern,
Quantity174 US, 37 OUS
DistributionDomestic distribution to CT, FL, NJ, SC, TN. Foreign distribution to Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211289RMU-2000 Automated Chest Compression System
Defibtech, LLC, Nov 9, 2021