Recall Z-2614-2024 ยท BTO Tube, Tracheostomy (W/Wo Connector)

Smiths Medical ASD Inc.

a) PEDIATRIC TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT NECK , Product Code/List Number/Item Code 67PFSS25; b) PEDIATRIC TRACHEOSTOMY TUBE 3.0mm TTS FLEXTEND STRAIGHT NECK , Product Code/List Number/Item Code 67PFSS30; c) PEDIATRIC TRACHEOSTOMY TUBE 3.5mm TTS FLEXTEND STRAIGHT NECK , Product Code/List Number/Item Code 67PFSS35; d) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND STRAIGHT NECK

DateAug 23, 2024 (initiated May 29, 2024)
StatusOpen, Classified
ReasonSmiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Root causeUnder Investigation by firm
ActionSmiths Medical issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 05/29/2024 via letter. The notice explained the issue with the device, potential risk and requested the following: "Customer Required Actions: When using the device, all instructions, including warning and cautions contained in the Instructions for Use Documentation must be followed with heightened awareness. Please complete the following actions listed below: 1. Check all inventory locations within your institution for the affected catalog numbers listed in the notification and discontinue use. Discard all a
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912469BIVONA WIRE REINFORCED PEDIATRIC TRACHEOSTOMY TUBE
Bivona Medical Technologies, Jun 19, 1991