ClearanceBrief
Recall Z-2608-2024 ยท BTO Tube, Tracheostomy (W/Wo Connector)

Smiths Medical ASD Inc.

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP25; b) NEONATAL TRACHEOSTOMY TUBE 3.0mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP30; c) NEONATAL TRACHEOSTOMY TUBE 3.5mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP35; d) NEONATAL TRACHEOSTOMY TUBE

DateAug 23, 2024 (initiated May 29, 2024)
StatusOpen, Classified
ReasonSmiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Root causeUnder Investigation by firm
ActionSmiths Medical issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 05/29/2024 via letter. The notice explained the issue with the device, potential risk and requested the following: "Customer Required Actions: When using the device, all instructions, including warning and cautions contained in the Instructions for Use Documentation must be followed with heightened awareness. Please complete the following actions listed below: 1. Check all inventory locations within your institution for the affected catalog numbers listed in the notification and discontinue use. Discard all a
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912469BIVONA WIRE REINFORCED PEDIATRIC TRACHEOSTOMY TUBE
Bivona Medical Technologies, Jun 19, 1991