Recall Z-2441-2024 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Sensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. Equipped by modules, enabling various configurations. Model: 11007641

DateJul 25, 2024 (initiated Jul 3, 2024)
StatusOpen, Classified
ReasonIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
Root causeSoftware design
ActionSiemens issued Urgent Medical Deice Correction letter (AX038/23/S-C) on 7/3/24 to Director of Diagnostic Imaging and/or Radiology Laboratory. Letter states reason for recall, health risk and action to take: If the Sensis documentation functionality is used and the described scenario occurs, users should continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. You can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. Please ensure that patient treatment can be con
Quantity12 units
DistributionUS Nationwide distribution in the states of CA, CT, FL, IA, KS, MA, MI, NJ, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223812Sensis Vibe (VD15)
Siemens Medical Solutions USA, Inc., Sep 15, 2023