ClearanceBrief
Recall Z-2564-2024 ยท OEZ Cardiovascular Procedure Kit

Merit Medical Systems, Inc.

Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K09/K09T (Custom Manifold Kit), REF: K09-00526L K09-00807AH K09-01467AD K09-01467AF K09-01514U K09-01734L K09-02040C K09-02683F K09-02804B K09-03844F K09-04088H K09-04088J K09-04863J K09-04946B K09-04998H K09-04998HP K09-05229JP K09-05342A K09-05375F K09-05939D K09-07759J K09-07789L K

DateAug 12, 2024 (initiated Jun 11, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeNo Marketing Application
ActionOn 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the
Quantity188837 kits
DistributionU.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.