ClearanceBrief
Recall Z-2562-2024 ยท OEQ Angiography/Angioplasty Kit

Merit Medical Systems, Inc.

Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF: K05-00203M K05-00511J K05-00547J K05-00612Q K05-00819A K05-01030H K05-01274D K05-01646 K05-01688A K05-01717B K05-01880K K05-01890D K05-01945A K05-01954A K05-02247A K05-02436B K05-02507 K05-02511A K05-02954 K05-02982A To support various vascular or cardi

DateAug 12, 2024 (initiated Jun 11, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeNo Marketing Application
ActionOn 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the
Quantity13,560 kits
DistributionU.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall