ClearanceBrief
Recall Z-2687-2024 ยท NDP Accessories, Implant, Dental, Endosseous

Preat Corp

Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203

DateAug 19, 2024 (initiated Jun 25, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.
Root causeProcess control
ActionOn 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email jespino@younginnovations.com
Quantity185 devices
DistributionU.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.