ClearanceBrief
Recall Z-3045-2024 ยท MLT Pelvic Exam Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline Convenience kits labeled as: 1) GROTH PELVISCOPY CDS, Pack Number CDS984187T ; 2) GROTH PELVISCOPY, Pack Number CDS984187U ; 3) DERMATOLOGY KIT, Pack Number DYNDA1567D; 4) ER MINOR SUTURE SET, Pack Number DYNDA1977A; 5) HSG TRAY, Pack Number DYNDH1453; 6) TRAY, ER LACERATION, Pack Number DYNDL1569A; 7) LAP/PELVISCOPY CSTM PK-LF , Pack Number DYNJ0157511K ; 8) SMSC-PELVISCOPY PACK-

DateSep 6, 2024 (initiated Apr 8, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeUnder Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pr
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.