Recall Z-2436-2024 ยท FRN Pump, Infusion

B Braun Medical Inc

Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A

DateJul 25, 2024 (initiated Jun 14, 2024)
StatusOpen, Classified
ReasonPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Root causeDevice Design
ActionOn June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.
Quantity930
DistributionDomestic distribution nationwide. International distribution to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092313PERFUSOR SPACE INFUSION SYRINGE PUMP SYSTEM
B.Braun Medical, Inc., Nov 4, 2009