ClearanceBrief
Recall Z-2433-2024 · FRN Pump, Infusion

B Braun Medical Inc

Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A

DateJul 25, 2024 (initiated Jun 14, 2024)
StatusOpen, Classified
ReasonPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Root causeDevice Design
ActionOn June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.
Quantity14949
DistributionDomestic distribution nationwide. International distribution to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092313PERFUSOR SPACE INFUSION SYRINGE PUMP SYSTEM
B.Braun Medical, Inc., Nov 4, 2009
K172831Perfusor® Space Syringe Infusion Pump System
B.Braun Medical, Inc., Jun 12, 2018
K191910SpaceStation MRI
B. Braun Melsugen AG, Mar 12, 2020