Recall Z-2997-2024 ยท OFF Central Venous Catheter Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline Convenience kits labeled as: 1) PUNCTURE TRAY , Pack Number 00-401649G ; 2) PICC LINE TRAY, Pack Number 00-401993O ; 3) ARTERIOGRAM TRAY, Pack Number 00-402001T ; 4) ARTERIOGRAM TRAY, Pack Number 00-402001U ; 5) HEALTH LINE MAX BARRIER KIT , Pack Number A134-MAX ; 6) ART-LINE SURGERY PACK-LF, Pack Number ARTNC175 ; 7) AV FISTULA PACK , Pack Number CDS840188M ; 8) VP SHUNT, Pack Nu
| Date | Sep 6, 2024 (initiated Apr 8, 2024) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pr |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall