ClearanceBrief
Recall Z-2982-2024 ยท CAE Airway, Oropharyngeal, Anesthesiology

MEDLINE INDUSTRIES, LP - Northfield

Medline Convenience kits, labeled as: 1) TRACH CDS-LF , Pack Number CDS983142B ; 2) MER TRACH CDS , Pack Number CDS984376B ; 3) Flexible LMA Pack with Syringe and Lube, Size 1, Pack Number DYND300010P; 4) Flexible LMA Pack with Syringe and Lube, Size 1.5, Pack Number DYND300015P; 5) Flexible LMA Pack with Syringe and Lube, Size 2, Pack Number DYND300020P; 6) Flexible LMA Pack with Syringe

DateSep 6, 2024 (initiated Apr 8, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeUnder Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pr
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.