Recall Z-2415-2024 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Olympus Corporation of the Americas

ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and endoscopic surgeries Model/Catalog Number: WA91307C

DateJul 23, 2024 (initiated Jun 21, 2024)
StatusOpen, Classified
ReasonElectrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
Root causeComponent design/selection
ActionOlympus issued URGENT: MEDICAL DEVICE REMOVAL Letter on 6/21/24. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to coordinate a repair to replace the power supply on your affected ESG-410 unit(s). Technical analysis of this issue performed by Olympus has determined that when the device is successfully powered on, it will work within specification. Therefore, you may continue to use your ESG-410 generator until the power supply is replaced. To minimize procedural delays, power on the generator prior to starting the procedure and,
Quantity86
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203277Electrosurgical Generator ESG-410, Foot Switches, PK Cutting Forceps
Olympus Winter & Ibe GmbH, Apr 28, 2021