Recall Z-2457-2024 ยท DWJ System, Thermal Regulating
Medivance Inc.
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun Stat Japan/600020; Arctic Sun 5000, NA/5000-00-00; Arctic Sun 5000, NA w/ Transmission Interface Module/5000-00-00E; Loaner, Arctic Sun M5000, NA/5000-00-00L; Arctic Sun 5000, EU/5000-1-01; Loaner, Arct
| Date | Jul 31, 2024 (initiated Jun 27, 2024) |
| Status | Open, Classified |
| Reason | If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue. |
| Root cause | Software design |
| Action | On 6/27/24, correction notices were sent to customers and distributors informing them of the following: Advice for Clinical Users: 1) Systems contains the following two fixed alarms when the device is operating in patient control mode: - Alarm 10 will alert if the patient s temperature and/or water temperature is below 31 degrees Celsius and the water temperature is below 31 degrees Celsius. - Alarm 08 will alert if the patient s temperature and/or water temperature is above 39.5 degrees Celsius. 2) Customers should continue to use the customizable Patient Temperature Low (Alarm 11) and Patie |
| Quantity | 10507 |
| Distribution | Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries o |