ClearanceBrief
Recall Z-2372-2024 ยท KWL Prosthesis, Hip, Hemi-, Femoral, Metal

Howmedica Osteonics Corp.

UNITRAX Endoprosthesis Head Component - 38mm. Implantable component used in partial hip arthroplasty. Catalog Number: 6942-5-038

DateJul 16, 2024 (initiated Jun 19, 2024)
StatusOpen, Classified
ReasonMislabeled: the size on the package label of the UNITRAX Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
Root causeEmployee error
ActionStryker issued Urgent Medical Device Recall (UMDR) PFA 3585464 third party Sedgwick on June 19, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected produ
Quantity12 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Hong Kong, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K014226UNITRAX MODULAR UNIPOLAR SYSTEM
Howmedica Osteonics Corp., Jan 25, 2002