Recall Z-2166-2024 ยท REM Spectroscopy Instrument

Horiba Instruments Incorporated

HORIBA custom configured fluorescence instrument, modular Fluorolog-QM

DateJul 1, 2024 (initiated Mar 28, 2024)
StatusOpen, Classified
ReasonHORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
Root causeRadiation Control for Health and Safety Act
ActionHoriba plans to issue an IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING letter to notify affected customers. The notification will Identify the affected product involved, description of the defect, safety hazard imposed: and corrective actions, For questions, call (732) 623-8146.
Quantity4 units
DistributionUS Nationwide and Worldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.