Recall Z-2248-2024 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Megadyne Medical Products, Inc.

Brand Name: MEGADYNE" MEGA 2000" Patient Return Electrode Product Name: MEGADYNE" MEGA 2000" Patient Return Electrode Model/Catalog Number: 0800 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega 2000, Adult, For Patients Weighing over 25lbs. (11.4Kg), 91cm Long x 51cm Wide Component: No

DateJul 3, 2024 (initiated Jun 17, 2024)
StatusOpen, Classified
ReasonTo limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
Root causeUnknown/Undetermined by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION letter, dated 6/17/24, was mailed to consignees. Consignees are instructed to share the recall notification with all users of affected devices and confirm their understanding, post a copy of the communication for user awareness, and forward the recall notification to any facility to which product was shipped. The notification reads that Mega Soft and Mega 2000 devices are only to be used for patients 12 years old. The notification also reminds consignees to follow proper cleaning, placement, and setup steps. Consignees are asked to complete and return the pr
Quantity259 units
DistributionDomestic: Nationwide distribution International: AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHINA, GERMANY, GREECE, INDIA, ITALY, NETHERLANDS, NORWAY, JAPAN, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, THAILAND, TURKEY, UNITED KINGDOM, UNITED STATES,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080741MEGA SOFT REUSABLE PATIENT RETURN ELECTRODE
Megadyne Medical Products, Inc., Dec 16, 2008