Recall Z-2350-2024 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Boston Scientific Corporation

Autotome RX 20mm CUT WIRE Cannulating Sphincterotome, Material Number (UPN) M00545170. The device is used in selective cannulation of the bile duct.

DateJul 12, 2024 (initiated Jun 12, 2024)
StatusOpen, Classified
ReasonForeign material may be present on the enclosed section of the cutting wire which may lead to inability to optimally bow the sphincterotome. This issue may lead to prolongation of procedure.
Root causeProcess control
ActionAll customers notified via courier on June 12, 2024 - notice was titled "Urgent Medical Device Removal." Customers were informed to cease further use or distribution of any remaining affected product. Product should be segregated and returned to the firm. For more information, please reach out to your local Boston Scientific representative.
Quantity256 US, 39 OUS
DistributionDomestic distribution nationwide.
FDA recordOpen on fda.gov