ClearanceBrief
Recall Z-2382-2024 ยท FPA Set, Administration, Intravascular

B. Braun Medical, Inc.

Infusomat SPACE PUMP SET 15 DROP W/2 SAFEDAY-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362420

DateJul 17, 2024 (initiated Jun 17, 2024)
StatusOpen, Classified
ReasonPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
Root causeProcess control
ActionB. Braun Medical Inc. (BBMI) is issuing a voluntary Urgent Medical Device Correction letter via US Postal Service Certified Mail with registered return receipt on 6/17/24. Letter states reason for recall, health risk and action to take: Corrective actions are being implemented to prevent these concerns in future manufacturing lots through optimization of the disks that provide the sealing function within these valves. Until such time as lots with these corrective actions are available, users are asked to follow the "Interim Measures for Users." A follow-up communication will be provided to use
Quantity431999 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142036INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP ADMINISTRATION SETS
B.Braun Medical, Inc., Aug 25, 2014