Recall Z-2411-2024 · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Waldemar Link GmbH & Co. KG (Mfg Site)
Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm
| Date | Jul 18, 2024 (initiated May 22, 2024) |
| Status | Open, Classified |
| Reason | Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique |
| Root cause | Process design |
| Action | LinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Pro |
| Quantity | 114 units US; 2351 units OUS |
| Distribution | Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slov |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K143179 | The LINK® Endo-Model® Knee System Waldemar Link GmbH & Co. KG, Jan 30, 2015 |
| K151008 | LINK® MEGASYSTEM-C® Waldemar Link GmbH & Co. KG, Aug 26, 2015 |
| K212742 | GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee System with LINK PorEx (TiNbN) inhouse coating Waldemar Link GmbH & Co. KG, Sep 16, 2021 |