Recall Z-2410-2024 · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Waldemar Link GmbH & Co. KG (Mfg Site)

Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-2817/11, 15-2817/12, 15-2818/11, 15-2818/12, 15-2834/01, 15-2834/02, 15-2834/03, 15-2834/04, 15-2835/11, 15-2835/12, 15-2836/11, 15-2836/12, 15-2837/11, 15-2837/12, 15-2838/11, 15-2838/12, 15-3815/11, 15-3815/12, 15-3816/1

DateJul 18, 2024 (initiated May 22, 2024)
StatusOpen, Classified
ReasonBlind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
Root causeProcess design
ActionLinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Pro
Quantity552 units; 7351 units OUS
DistributionNationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slov
FDA recordOpen on fda.gov