ClearanceBrief
Recall Z-2429-2024 ยท MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor

Abbott Medical

Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only

DateJul 25, 2024 (initiated May 22, 2024)
StatusOpen, Classified
ReasonTheir is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
Root causeError in labeling
ActionOn June 6, 2024, Abbott, issued a "Urgent Medical Device Correction" notification to affected consignees via mail. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. For information on when an ERI notification appears and how to make a determination with your patients on timing of IPG replacement, see the customer communication. 3. For recommended strategies for preventing lapse in therapy, see the customer com
Quantity5,900 units
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.