Recall Z-2428-2024 ยท CFN Method, Nephelometric, Immunoglobulins (G, A, M)

Beckman Coulter Inc.

Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

DateJul 24, 2024 (initiated May 27, 2024)
StatusOpen, Classified
ReasonLIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
Root causeProcess control
ActionOn May 23, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification via: xxx. 1. Update the LIH Influence Check settings for Lipemia on AU480, AU680, DxC 500 AU and DxC 700 AU analyzers if they are enabled for IgM. 2. To determine if the LIH Influence Check is enabled for IgM and to update your Lipemia setting, see customer communication. 3. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. . Discontinuance or disposal of this product is
Quantity13,565 units
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Albania, Algeria, Argentina, Australia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall