ClearanceBrief
Recall Z-2165-2024 ยท MOS Coil, Magnetic Resonance, Specialty

Philips North America Llc

SENSE XL TORSO COIL 1.5T Mk2. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567502281, 453567502282, 453567502283, 453567502284, 453567502285, 453567502286, 459801585212.

DateJun 27, 2024 (initiated Jun 5, 2024)
StatusOpen, Classified
ReasonPotential for coils to heat up and harm patients (burn).
Root causeDevice Design
ActionAn URGENT MEDICAL DEVICE RECALL notice, dated 5/31/24, was mailed to consignees. The notice instructs users to follow the provided instructions when operating a MR system with affected coils. These instructions include avoiding First Level Operating Mode/High SAR scans, use dedicated pads and mattresses provided with the coils. avoid positioning the coil closer than 2 inches (5 cm) from the bore, and do not exceed 45 minutes of examination time. An Advisory Notice that summarizes the recommended actions was provided with the recall notification and is to be displayed with affected systems to e
Quantity14 units
DistributionWorldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Ko
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043147ACHIEVA FAMILY
Philips Medical Systems North America Co., Feb 15, 2005
K052013ACHIEVA 1.5T & INERA 1.5T FAMILY
Philips Medical Systems Nederland B.V., Aug 2, 2005