Recall Z-2313-2024 ยท KQM Camera, Surgical And Accessories

Baxter Healthcare Corporation

VidiaPort carrying arm, mount for auxiliary products, model numbers: a) 4028110 (Ceiling Single Surgical Light); b) 4028210 (Ceiling Duo Surgical Lights); c) 4028310 (Ceiling Trio Surgical Lights)

DateJul 5, 2024 (initiated Jun 4, 2024)
StatusOpen, Classified
ReasonBaxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
Root causeUnder Investigation by firm
ActionBaxter issued an Urgent Medical Device Correction notice to its consignees on 06/04/2024 via US mail, first class. The notice explained the issue with the devices, potential hazards involved, and requested the following: "Actions to be Taken by Customers: 1. Operators may continue to use the iLED 7 surgical light systems by following the current instructions for use (IFU) cautions and warnings while considering the additional information below. . Utilize the lowest possible illumination level suitable for the procedure, especially in certain neurological or intestinal procedures on delica
Quantity275 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.