Recall Z-1946-2024 ยท RFN Fiber Optic Communication And Data Transfer

Newport Corp

Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.

DateJun 5, 2024 (initiated Mar 27, 2024)
StatusOpen, Classified
ReasonAccording to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Root causeRadiation Control for Health and Safety Act
ActionNewport will contact each affected customer, distributor and end user, to determine the status of the laser s current hardware and whether the power supply must be replaced. If the power supply has the incorrect hardware and must be replaced, Newport will remedy the defect by replacing the unit with a compliant unit.
Quantity19
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.