Recall Z-2746-2024 ยท OFG Ct Biopsy Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline procedural kits labeled as: 1) BIOPSY PACK, Pack Number DYNJ63370A; 2) CIRCUMCISION TRAY, Pack Number CIT7210; 3) CIRCUMCISION TRAY, Pack Number DYNDA2271; 4) MERCY MEDICAL CIRCUMCISION TRAY, Pack Number CIT6780; 5) UCLA - BREAST BIOPSY KITDYNDH1512A
| Date | Aug 28, 2024 (initiated Apr 8, 2024) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pro |
| Distribution | Worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.