Recall Z-2220-2024 ยท OZE Influenza A And Influenza B Multiplex Nucleic Acid Assay

Centers For Disease Control and Prevention

CDC, Influenza A/H5 Subtyping Kit, Model/Catalog Number: FluIVD03-11

DateJun 27, 2024 (initiated Apr 26, 2024)
StatusTerminated
ReasonThe H5b component may fail to amplify, resulting in an inconclusive result.
Root causeUnder Investigation by firm
ActionThe CDC notified customers via letter on 04/26/2024. Customers were instructed to halt use of the H5b component when testing a specimen with the CDC Influenza A/H5 subtyping kit, lot 220307. Proceed with testing with the InfA, H5a, and RP targets and do not delay resulting. If a specimen is positive for InfA and H5a, that is considered a presumptive H5 positive and should be sent to CDC immediately. Please reach out to flusupport@cdc.gov for further shipping guidance. If a specimen is positive for InfA and negative for H5a and has not been tested with the Influenza A Subtyping kit, procee
Quantity191 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200370CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A/B Typing Kit, CDC Human Influenza Virus Real-ti
Centers for Disease Control and Prevention, Mar 10, 2020