Recall Z-2332-2024 ยท DXJ Display, Cathode-Ray Tube, Medical

Barco N.V.

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

DateJul 11, 2024 (initiated Apr 24, 2024)
StatusCompleted
ReasonWe have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
Root causeOther
ActionBarco NV notified its sole consignee on 04/29/2024 via email. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by the customer: 1. Identify the location and amount of all Molex modules that are listed in pdf document Molex_SN_Partner . 2. Fill out and send back Response Form 1 to your Barco Key Account Manager, including the number of distributed units per country. 3. Disable or discard all used Molex modules. 4. Fill out and send back Response Form 2 to your Barco Key Account Manager after completion of the actions. For question, contact yo
Quantity88 units
DistributionUS: FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173381Nexxis OR
Barco N.V., Feb 22, 2018