Recall Z-2219-2024 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Stryker GmbH

Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.

DateJun 27, 2024 (initiated May 9, 2024)
StatusOpen, Classified
ReasonThe adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
Root causeProcess change control
ActionThe recalling firm issued letters dated 5/9/2024 on 5/9/2024 and 5/11/2024 via email to sales reps and distributors and via FedEx ground to hospitals. The letter described the product, the issue, and potential risks. The consignees instructions were: (1) Immediately check your inventory to locate the affected product listed on the Business Reply Form, remove the product from its point of use, and isolate it to prevent accidental use; (2) Sign and return the enclosed Business Reply Form by email by 5/15/2024 to confirm receipt of the notification/documenting product disposition: (3) Upon rec
Quantity39 devices
DistributionWorldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, KY, MA, MI, NE, NJ, OH, OR, and TX. The countries of Germany, Japan, Netherlands, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.