Recall Z-2233-2024 ยท NAY System, Surgical, Computer Controlled Instrument

Olympus Corporation of the Americas

EasySuite 4K Integrated Operating Room System Model/Catalog Number: EasySuite 4K- The EasySuite 4K system includes a touch panel user interface for quick routing of 2K/4K procedural video to displays within the operating room and for convenient control of room video cameras, surgical cameras, ambient lighting, and surgical lights.

DateJun 28, 2024 (initiated Apr 24, 2024)
StatusOpen, Classified
ReasonA subset of the ES4K systems do not have complete records for earth leakage testing. The earth leakage test is a series of measurements intended to confirm that the amount of electrical current flowing to the earth is within standard limits
Root causeProcess control
ActionOlympus issued Urgent: Medical Device Corrective Action letter on May 24, 2024 via Courier, 2-day delivery. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to schedule an on-site service to perform the earth leakage testing on your facility s ES4K system. You may continue to use your ES4K system until an Olympus representative performs the earth leakage test. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, ident
Quantity227 units
DistributionWorldwide distribution - US Nationwide and the countries of Germany, Hong Kong.