Recall Z-2212-2024 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)
Medtronic Inc.
Product not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model Number DTBA2QQ. b. CLARIA MRI CRT-D SureScan, Model Number DTMA2D1. c. CLARIA MRI QUAD CRT-D SureScan, Model Number DTMA2QQ. d. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D1. e. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D4. f. AMPLIA MRI QUAD CRT-D SureScan, Model Number DTMB2
| Date | Jun 26, 2024 (initiated Apr 16, 2024) |
| Status | Open, Classified |
| Reason | Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing. |
| Root cause | Nonconforming Material/Component |
| Action | On about 04/18/2024, Medtronic notified customers that they were retrieving distributed units that may have undergone a specific manufacturing sequence that requires additional engineering evaluation. They requested that customers acknowledge the letter with signature. |
| Quantity | 82 units |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Mol |
| FDA record | Open on fda.gov |