Recall Z-2773-2024 ยท OHG Circumcision Tray
MEDLINE INDUSTRIES, LP - Northfield
Centurion procedure packs, labeled as: 1) BIOPSY TRAY-LF, Pack Number SPEC4380; 2) CIRCUMCISION TRAY, Pack Number CIT6325; 3) CIRCUMCISION/MINOR TRAY, Pack Number CIT6785A; 4) LUMBAR PUNCTURE TRAY, Pack Number PT215; 5) ULTRASOUND GUIDED PIV KIT, Pack Number MNS9580; 6) VANTEX 7FR 3L 16CM CVC BUNDLE US, Pack Number ECVC8050
| Date | Aug 28, 2024 (initiated Apr 8, 2024) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request that the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other |
| Distribution | worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.