Recall Z-2140-2024 · CBK Ventilator, Continuous, Facility Use

Baxter Healthcare Corporation

Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.

DateJun 26, 2024 (initiated May 29, 2024)
StatusOpen, Classified
ReasonPotential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.
Root causeUnder Investigation by firm
ActionBaxter Healthcare began notifying consignees on about 05/29/2024 via " URGENT MEDICAL DEVICE RECALL" letter sent U.S.P.S first class mail. Patients, healthcare providers, wholesalers and distributors were instructed to ensure patients always have an alternate means of ventilator or oxygen therapy available, inspect the battery charger dongle for damage and replace immediately if damaged, and patients may continue to use the ventilator system once they inspect and confirm no damage to the dongle and battery is charging appropriately. Patients were also asked to observe any alarms of the venti
Quantity2,510 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170037Breathe Technologies Life2000™ Ventilation System