Recall Z-2444-2024 ยท OES Cardiac Catheterization Kit
MEDLINE INDUSTRIES, LP - Northfield
Medline procedure kits labeled as: 1) WMC HYBRID OR PACK PART 1 , Pack Number CMPJ08068F ; 2) VASCULAR PACK , Pack Number CMPJ11455B ; 3) SALEM CATH LAB-LF , Pack Number DYNJ26938J ; 4) EH PEDIATRIC CATH , Pack Number DYNJ49920L ; 5) ARTERIOGRAM PACK, Pack Number DYNJ53018; 6) CATH PACK , Pack Number DYNJ57729C ; 7) PEDS CATH PACK, Pack Number DYNJ66856; 8) PEDS CATH PACK, Pack Number DYN
| Date | Aug 8, 2024 (initiated Apr 8, 2024) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pro |
| Quantity | 1,679,067 units in total |
| Distribution | Worldwide Distribution: US (nationwide) and OUS countries of: AE, AI, BB, CA, CL, CY, DE, GT, HN, JP, MX, MY, PA, PE, PR, SA, SG, TC, TT, VI, PR, TT, MX, BM, ES, NI, DO, GU, CR, AT, CD, DE, DF, W, and ON. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.