ClearanceBrief
Recall Z-2623-2024 ยท OFQ Anesthesia Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline procedure packs containing Shenli syringes, labeled as follows: 1) LOCAL PACK, pack number DYNJ34129C, 2) FTMC LOCAL PACK-LF, pack number DYNJ54607A, 3) LOCAL PACK, pack number DYNJ61839B

DateAug 27, 2024 (initiated Apr 8, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeNonconforming Material/Component
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pr
Quantity1,679,067 units in total
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.