ClearanceBrief
Recall Z-2318-2024 ยท NHA Abutment, Implant, Dental, Endosseous

Preat Corp

9006767 Astra EV 4.8mm Milled Titanium Abutment Milled abutments are too be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.

DateJul 8, 2024 (initiated Apr 10, 2024)
StatusOpen, Classified
ReasonDue to incorrect milled titanium abutment being packaged and labeled.
Root causeProcess control
ActionOn 05/21/2024, the firm sent via direct certified mail an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter to customers informing them that product labeled as part number 9006767-F Astra EV 4.8mm Milled Titanium Abutment may contain instead part number 9006667 Astra EV 4.2mm Milled Titanium Abutment. The potential risk identified was, the screw seat is higher than specified in the engineering drawings and required for full thread engagement with the implant. This condition reduces the screw thread engagement with the implant. Customer are instructed to: Product in the scope of this recall shall
Quantity7 abutmens
DistributionU.S.: CA, MO, and NV. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220823Preat Abutments
Preat Corporation, Oct 26, 2022